流速测量 |
Method | |
量程 | | 0.1 ml/h to 1500 ml/h (2600 m/h is shown) |
|
精度 | | 1 % of reading ±1 LSD for flows of 16 to 200 ml/h for volumes over 20 ml, otherwise 2 % of reading ±1 LSD for volumes over 10 ml under laboratory conditions.Degassed water at 15 °C to 30 °C (59 °F to 86 °F) is recommended for long tests. |
|
zui长测试持续时间 | |
容积测量 |
Method | | 由测量模块直接测得容积,zui小采样量为 60 微升 |
|
量程 | |
精度 | | 流速在 16 ml/h 至 200 ml/h ,且容积大于 20 ml 时,为读数的 1 % ± 1 LSD。否则在实验室条件下大于 10 ml 时,为读数的 2 % ± 1 LSD。 |
|
zui长测试持续时间 | |
PCA bolus/dual flow measurement |
Method | | See volume measurement above |
|
Min bolus volume | |
分辨率 | |
Max test durations | |
压力测量 |
Method (back pressure and flow test) | | See volume measurement above |
|
量程 | | 0 psi 到 45 psi 或等同于 mmHg 和 kPa |
|
精度 | |
zui长测试持续时间 | |
Other specification |
模板 | | Predetermined test sequences.Typical capacity 200. |
|
结果存储 | | Test results stored for later viewing, printing or transfer to PC. Typical capacity 250 tests. |
|
一般规格 |
Operating voltage range | |
Supply frequency | |
Supply power | |
Fuses | |
大小 (高 x 宽 x 深) | | 30 cm x 20 cm x 20 cm | | (12 in x 8 in x 8 in) |
|
重量 | |
海拔高度 | | 0 m 至 3000 m (0 ft 至 10000 ft) |
|
温度 |
操作 | | 15 °C 至 30 °C(59 °F 至 86 °F) |
|
存储 | | 所有液体排出时,-20 °C 至 +40 °C (-4 °F 至 +104 °F) |
|
湿度 | |