StratAltitudeChambersT5ST/10ST/15ST/27ST/36ST/48ST/64STTenney’sTenneyStratAltitudeChamberssimulatealtitudesupto200,000feet
Strat Altitude Chambers T5ST/10ST/15ST/27ST/36ST/48ST/64ST |
Tenney’s Tenney Strat Altitude Chambers simulate altitudes up to 200,000 feet. With an overall temperature range of -70°C to +177°C, humidity capabilities of 20% to 98% RH, and workspace sizes ranging from 5- to 64-cubic-feet, this highly developed line of altitude chambers offers one of the broadest selections in the industry.
We are confident that our solid altitude chamber construction and precise operating systems will provide the exacting performance and a high degree of reliability for your altitude testing needs.
Strat Altitude Chambers T5ST/10ST/15ST/27ST/36ST/48ST/64ST Features: |
Continuously welded, stainless steel, vapor-tight interior liners |
Control tolerance is ±0.3°C and ±2% RH typical after stabilization |
Linear pressure transducer to measure altitude |
Non-corroding humidity system |
Low-water protection on humidity system |
Vertical-down recirculating conditioning stream |
Strat Altitude Chambers Specifications: |
Model | T5ST | 10ST | 15ST | 27ST | 36ST | 48ST | 64ST | |||||||||||
System Horsepower Size | 4 | 6 | 15 | 6 | 15 | 6 | 15 | 20 | 6 | 15 | 20 | 15 | 20 | 30 | 15 | 20 | 30 | |
Workspace inches (mm) | W | 21/53 (533/1346) | 24/61 (610/1549) | 24/61 (609/1549) | 36/91 (914/2311) | 36/91 (914/2311) | 48/122 (1219/3099) | 48/122 (1219/3099) | ||||||||||
D | 21/53 (533/1346) | 24/61 (610/1549) | 36/91 (914/2311) | 36/91 (914/2311) | 48/122 (1219/3099) | 36/91 (914/2311) | 48/122 (1219/3099) | |||||||||||
H | 25/51 (635/1295) | 30/76 (762/1930) | 30/76 (762/1930) | 36/91 (914/2311) | 36/91 (914/2311) | 48/122 (1219/3099) | 48/122 (1219/3099) | |||||||||||
Exterior inches (mm) | W | 25/51 (635/1295) | 51/130 (1295/3302) | 52/130 (1320/3302) | 63/160 (1600/4064) | 63/160 (1600/4064) | 77/196 (1956/4978) | 77/196 (1956/4978) | ||||||||||
D | 63/152 (1600/3860) | 99/203 (2515/5156) | 111/236 (2819/5994) | 111/241 (2819/6121) | 123/272 (3124/6909) | 123½ / 281 (3131/7137) | 135½ / 312 (3435/7924) | |||||||||||
H | 78/183 (1981/4648) | 84/210 (2134/5334) | 84/213 (2134/5410) | 90/229 (2286/5817) | 90/229 (2286/5817) | 91/231 (2311/5867) | 91/231 (2311/5867) | |||||||||||
Utilities, etc . | ||||||||||||||||||
Heater Capacity | 2 | 4 | 6 | 8 | 8 | 16 | 16 | |||||||||||
Humidifier | Watts | 1 | 2 | 3 | ||||||||||||||
GPH | 1 .05 | 1 .0 | 1 .5 | |||||||||||||||
AMPS @ 230V, 33 | 33(1) | 48 | 81 | 52 | 85 | 63 | 96 | 130 | 63 | 96 | 130 | 120 | 149 | 183 | 120 | 149 | 183 | |
AMPS, Fuse | 45 | 70 | 110 | 70 | 120 | 80 | 125 | 175 | 80 | 125 | 175 | 150 | 200 | 250 | 150 | 200 | 250 | |
Cooling Water, 20°C GPM | - | 3 | 6 | 3 | 6 | 3 | 6 | 9 | 3 | 6 | 9 | 6 | 9 | 12 | 6 | 9 | 12 | |
Unit Weight | Lbs/Kg | 1300/590 | 3150/1429 | 4850/2200 | 3150/1429 | 6550/2971 | 11,500/5216 | 13,050/5920 |